Chapter 2: Control Testing 16
|
|
|
|
|
|
|
|
|
|
|
|
Running a Control Test (continued)
6. Squeeze the bottle until a tiny drop forms at the
tip of the bottle. Touch the drop to the tip of the
yellow end of the test strip. Do not put control
solution on top of the test strip. When you
see flash, you have enough control solution
in the test strip. Wipe the tip of the bottle with a
tissue, then cap the bottle tightly.
7. A result appears on the display, along with a
control bottle symbol and a flashing “L.” Don’t
remove the test strip yet. Press once to
mark it as a Level 1. If you tested the Level 2
control, press a second time.