Accu-Chek Go Manuale di Servizio Pagina 178

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 208
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 177
178
Symbols, error messages and troubleshooting
Farbe: P 021 – Prüfmittelnummer: 201
This error message ... means:

O The battery is empty or a meter error
has occurred.
Remove the test strip by pushing the
ejector down and re-insert it. If the
error message persists, change the
battery.
O The meter was exposed to a strong
electromagnetic field. Eject the test
strip or turn the meter off. Move
somewhere else or turn off the
source of the field.
Vedere la pagina 177