Symbols, error messages and troubleshooting
Farbe: P 021 – Prüfmittelnummer: 201
Symbol What it means

Before or during a test or in memory: at
the time of testing the temperature is /
was outside the permitted range.
15
The beep tone (0n1) or Acoustic Mode
(0n2) is turned on.
16
or
The battery is running low.
When the symbol first appears in the dis-
play, you can perform about 50 more
tests.
When the battery is so low that no more
tests can be performed, the battery sym-
bol alone is displayed for 3 seconds. After
that, the meter turns off. Insert a new
battery.
17
When an average is displayed (from
memory): the symbol is followed by the
number of tests n included in the calcula-
tion.